Test-Setup

IEC 60601-1-2 Ed. 4.1
IEC 61000-4-39

IEC 60601-1-2 is the international standard for testing medical equipment for electromagnetic compatibility (EMC). The latest update adds a new test due to more potential sources of interference in homes, offices (places where medical equipment may be used), and hospitals. Three common frequencies are: 30 kHz, 134.2 kHz and 13.56 MHz. For this test, reference is made to the IEC 61000-4-39 standard for testing of fields in close proximity. The IEC 60601-1-2 test setup can be used to generate these test fields.

 

Performance features

Brief description
The test equipment includes the CDG 7000-MGA - 10 kHz - 250 MHz, 75 watt generator with amplifier, RF generator and RF meter x3 (monitored level, forward and reverse power). The system includes all cabling and HELIA 7 software necessary to meet the requirements of the standard.

Depending on the frequency range, the following complete set is required:

Required complete set for frequency range 9 kHz - 150 kHz:

  • MGA RL 120 – Radiating loop 120 mm as specified in IEC 61000-4-39 for 9 kHz-150 kHz, IEC / EN 60601-1-2 and MIL-STD-461 / RS101, 3 m cabl
  • MGA RL 120 NW – Matching Network for MGA RL 120, matches loop to 50 Ohms of the CDG 7000 amplifier acc. to IEC 61000-4-39 and IEC / EN 60601-1-2 for Immunity to magnetic fields 9 kHz - 150 kHz
  • MGA RL 120 Stand – Stand for MGA RL 120 for tests
  • MGA LS 040(additionally strongly recommended) – Loop sensor 40 mm that attaches to MGA RL 120 at the correct distance of 50 mm

Required complete set for frequency range 150 kHz - 26 MHz:

  • MGA RL 100-3– Radiating loop as specified in IEC 61000-4-39 and IEC / EN 60601-1-2 for 150 kHz – 26 MHz
  • MGA LS 040-1– Loop sensor 40 mm that attaches to MGA RL 100-3 at the correct distance of 50 mm as specified in IEC 61000-4-39 and IEC / EN 60601-1-2
  • MGA RL 100-3 stand – for tests in stronger magnetic field

In addition, one of the two adaptation networks is required:

  • MGA RL 100-3 NW-60601 – Matching Network 60601 for MGA RL 100-3, matches MGA RL 100-3 to the IEC 60601-1-2 requirements or
  • MGA RL 100-3 NW-61000 – Matching Network 61000 for MGA RL 100-3, matches MGA RL 100-3 to the IEC 61000-4-39 requirements